Endotracheal tubes and fluid aspiration: Evaluation of new c
Now open: Certificate Course in Management of Covid-19 by Govt. Of Gujarat and PlexusMDKnow more...Now open: Certificate Course in Management of Covid-19 by Govt. Of Gujarat and PlexusMDKnow more...
Abstract

Background :
Aspiration of subglottic secretions past the endotracheal tube (ETT) cuff is a prerequisite for developing ventilator-associated pneumonia (VAP). Subglottic secretion drainage (SSD) ETTs reduce aspiration of subglottic secretions and have demonstrated lower VAP rates. We compared the performance of seven SSD ETTs against a non-SSD ETT in preventing aspiration below inflated cuffs.

Methods :
ETTs were positioned vertically in 2 cm diameter cylinders. Four ml of a standard microbial suspension was added above inflated cuffs. After 1 h, aspiration was measured and ETTs demonstrating no leakage were subjected to rotational movement and evaluation over 24 h. Collected aspirated fluid was used to inoculate agar media and incubated aerobically at 37 °C for 24 h. The aspiration rate, volume and number of microorganisms that leaked past the cuff was measured. Experiments were repeated (×10) for each type of ETT, with new ETTs used for each repeat. Best performing ETTs were then tested in five different cylinder diameters (1.6, 1.8, 2.0, 2.2 and 2.4 cm). Experiments were repeated as above using sterile water. Volume and time taken for aspiration past the cuff was measured. Experiments were repeated (×10) for each type of ETT. Results were analysed using non-parametric tests for repeated measures......