This study assesses the reliability of successive corneal biomechanical response measurements by the Corneal Visualization Scheimpflug Technology in different keratoconus (KC) stages.

A total of 173 eyes were repeatedly examined five times with the CST, each after repositioning the patient’s head and re-adjusting the device. Tomographical analysis was performed once before and once after CST measurements.

Outcome measures included- A1 velocity,deformation amplitude (DA) ratio 2 mm,integrated radius, stiffness parameter A1 and Ambrosio relational thickness to the horizontal profile (ARTh). The Corvis Biomechanical Index (CBI) is reported to be extracted out of these parameters.

Results:
--Ambrosio relational thickness to the horizontal profile and SPA1 were significantly higher in controls (534|123) compared to TKC1 (384|88), TKC2 (232|66), TKC3 (152|55) and TKC4 (71|27).

--The other parameters were similar in controls and TKC1 (A1 velocity: 0.148|0.151 m/s; integrated radius: 8.2|8.6 mm-1), but significantly higher in TKC stages 2 to 4 (DA ratio 2 mm: 5.5|6.3|8.0; A1 velocity: 0.173|0.174|0.186 m/second; integrated radius: 10.9|12.8|19.0 mm-1).

--All parameters proved to be highly reliable (Cronbach’s alpha more than 0.834) and the corneal tomography remained unaffected.

Conclusively, the individual parameters included in the CBI (consisting of ARTh, SPA1, DA ratio 2 mm, A1 velocity and integrated radius) are highly reliable but differ KC stage dependently.